<?xml version="1.0" encoding="utf-8"?>
<!DOCTYPE article PUBLIC "-//NLM//DTD JATS (Z39.96) Journal Publishing DTD v1.0 20120330//EN" "JATS-journalpublishing1.dtd">
<article article-type="abstract" dtd-version="1.0" xml:lang="en" xmlns:xlink="http://www.w3.org/1999/xlink" xmlns:mml="http://www.w3.org/1998/Math/MathML">
<front>
<journal-meta>
<journal-id journal-id-type="publisher-id">CC</journal-id>
<journal-id journal-id-type="nlm-ta">Cardiol Croat</journal-id>
<journal-title-group>
<journal-title>Cardiologia Croatica</journal-title>
<abbrev-journal-title abbrev-type="pubmed">Cardiol. Croat.</abbrev-journal-title>
</journal-title-group>
<issn pub-type="ppub">1848-543X</issn>
<issn pub-type="epub">1848-5448</issn>
<publisher><publisher-name>Croatian Cardiac Society</publisher-name></publisher>
</journal-meta>
<article-meta>
<article-id pub-id-type="publisher-id">CC_2014_9_3-4_90</article-id>
<article-id pub-id-type="doi">10.15836/ccar.2014.90</article-id>
<article-categories><subj-group subj-group-type="heading"><subject>Extended Abstract</subject></subj-group>
</article-categories>
<title-group>
<article-title>Cardiovascular implantable electronic devices and electromagnetic interference</article-title>
</title-group>
<contrib-group>
<contrib contrib-type="author" corresp="yes"><name><surname>Vrazic</surname><given-names>Hrvoje</given-names></name><xref ref-type="aff" rid="aff1"><sup>1</sup></xref><xref ref-type="aff" rid="aff2"><sup>2</sup></xref></contrib><contrib contrib-type="author"><name><surname>Kolb</surname><given-names>Christof</given-names></name><xref ref-type="aff" rid="aff2"><sup>2</sup></xref></contrib>
<aff id="aff1"><label>1</label><institution>University Hospital Dubrava</institution>, <addr-line>Zagreb</addr-line>, <country country="hr">Croatia</country></aff>
<aff id="aff2"><label>2</label><institution>German Heart Centre of the State of Bavaria and the Technical University Munich</institution>, <addr-line>Munich</addr-line>, <country country="de">Germany</country></aff>
</contrib-group>
<author-notes>
<corresp id="cor1">Correspondence to Hrvoje Vrazic, Klinicka bolnica Dubrava, Avenija Gojka Suska 6, HR-10000 Zagreb, Croatia; Phone: +385-1-2902-444; E-mail: <email xlink:href="vrazic@gmail.com">vrazic@gmail.com</email></corresp></author-notes>
<pub-date date-type="pub" publication-format="electronic"><month>03</month><year>2014</year></pub-date>
<pub-date date-type="pub" publication-format="print"><month>03</month><year>2014</year></pub-date>
<volume>9</volume>
<issue>3-4</issue>
<fpage>90</fpage>
<lpage>90</lpage>
<permissions>
<copyright-statement>Croatian Cardiac Society</copyright-statement>
<copyright-year>2014</copyright-year>
<copyright-holder>Croatian Cardiac Society</copyright-holder>
</permissions>
<kwd-group kwd-group-type="author"><title>KEYWORDS: </title><kwd>cardiovascular implantable electronic device</kwd><kwd>pacemaker</kwd><kwd>implantable cardioverter defibrillator</kwd><kwd>electromagnetic interference</kwd></kwd-group>
</article-meta>
</front>
<body>
<p>Introduction: Most frequently used cardiovascular implantable electronic devices for the treatment of arrhythmias are pacemakers and implantable cardioverter defibrillators (ICDs). It has been shown that the function of these cardiac rhythm devices can be impaired by electromagnetic interference from the devices and systems emitting magnetic fields, causing either temporary or permanent system malfunction.</p>
<p>Methods: Literature review (<xref ref-type="bibr" rid="r1"><italic>1</italic></xref>-<xref ref-type="bibr" rid="r3"><italic>3</italic></xref>) and own results (<xref ref-type="bibr" rid="r4"><italic>4</italic></xref>) from the studies examining the potential electromagnetic interference of hand metal detectors among the patients with implanted cardiovascular implantable electronic devices performed in Deutsches Herzzentrum Munich will be used.</p>
<p>Results: Sources of electromagnetic interference are frequently encountered in daily life, emanating from cellular phones, anti-theft devices, metal detectors, various remote controls, improperly wired/grounded appliances coming into the contact with the body, MP3 players, induction ovens and many other devices (<xref ref-type="bibr" rid="r1"><italic>1</italic></xref>-<xref ref-type="bibr" rid="r3"><italic>3</italic></xref>) Increasing global efforts to intensify security screening measures in the past decade have also brought into focus the use of metal detectors (which create a magnetic field that might interfere with the function of cardiovascular implantable electronic devices); they are nowadays routinely used in many settings to detect ferrous and other dangerous items. Patients with pacemakers and implantable cardioverter defibrillators are often advised to avoid screening with metal detectors because of the risk for electromagnetic interference. Another important group of sources of electromagnetic interference is that which is encountered in workplace and/or industrial environments, such as high voltage power lines, transformers, welders, electric motors, and many others. Last, but not least, are those sources are found in the medical environment: magnetic resonance image scanners, electrosurgery, defibrillation, neurostimulators, TENS devices, radiofrequency catheter ablation, and therapeutic diathermy.</p>
<p>Conclusions: Altough the majority of data available is based on case reports, there are a few studies examining effects of electromagnetic intereference on cardiovascular implantable electronic devices. Majority of problems that occur are transient in nature, and, more important, easily avoidable. Physicians should be aware of and familiarized with the most common interactions of electromagnetic intereference and cardiovascular implantable electronic devices.</p>
</body>
<back>
<ref-list>
<title>Literature</title>
<ref id="r1"><label>1</label><mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Pinski</surname><given-names>SL</given-names></name><name><surname>Trohman</surname><given-names>RG</given-names></name></person-group>. <article-title>Interference in implanted cardiac devices, Part I.</article-title> <source>Pacing Clin Electrophysiol</source>. <year>2002</year>;<volume>25</volume>:<fpage>1367</fpage>&#x2013;<lpage>81</lpage>. <pub-id pub-id-type="doi">10.1046/j.1460-9592.2002.01367.x</pub-id><pub-id pub-id-type="pmid">12380774</pub-id></mixed-citation></ref>
<ref id="r2"><label>2</label><mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Thaker</surname><given-names>JP</given-names></name><name><surname>Patel</surname><given-names>MB</given-names></name><name><surname>Jongnarangsin</surname><given-names>K</given-names></name><name><surname>Liepa</surname><given-names>VV</given-names></name><name><surname>Thakur</surname><given-names>RK</given-names></name></person-group>. <article-title>Electromagnetic interference with pacemakers caused by portable media players.</article-title> <source>Heart Rhythm</source>. <year>2008</year>;<volume>5</volume>:<fpage>538</fpage>&#x2013;<lpage>44</lpage>. <pub-id pub-id-type="doi">10.1016/j.hrthm.2008.01.028</pub-id><pub-id pub-id-type="pmid">18329961</pub-id></mixed-citation></ref>
<ref id="r3"><label>3</label><mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Trigano</surname><given-names>A</given-names></name><name><surname>Blandeau</surname><given-names>O</given-names></name><name><surname>Dale</surname><given-names>C</given-names></name><name><surname>Wong</surname><given-names>MF</given-names></name><name><surname>Wiart</surname><given-names>J</given-names></name></person-group>. <article-title>Reliability of electromagnetic filters of cardiac pacemakers tested by cellular telephone ringing.</article-title> <source>Heart Rhythm</source>. <year>2005</year>;<volume>2</volume>:<fpage>837</fpage>&#x2013;<lpage>41</lpage>. <pub-id pub-id-type="doi">10.1016/j.hrthm.2005.03.011</pub-id><pub-id pub-id-type="pmid">16051120</pub-id></mixed-citation></ref>
<ref id="r4"><label>4</label><mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Jilek</surname><given-names>C</given-names></name><name><surname>Tzeis</surname><given-names>S</given-names></name><name><surname>Vrazic</surname><given-names>H</given-names></name><etal/></person-group> <article-title>Safety of screening procedures with hand-held metal detectors among patients with implanted cardiac rhythm devices: a cross-sectional analysis.</article-title> <source>Ann Intern Med</source>. <year>2011</year>;<volume>155</volume>(<issue>9</issue>):<fpage>587</fpage>&#x2013;<lpage>92</lpage>. <pub-id pub-id-type="doi">10.7326/0003-4819-155-9-201111010-00005</pub-id><pub-id pub-id-type="pmid">22041947</pub-id></mixed-citation></ref>
</ref-list>
</back>
</article>